LCR Meters - 4310 4320 4350 and 43100
|
|
|
|
- Comprehensive measurement functions, including DC resistance
(Optional)
- Test components from 20Hz to 1 MHz (43100)
- 0.1% basic accuracy
- Drive level from 10mV to 2Vrms
- Internal and external DC Voltage bias
- GPIB / IEEE-488 and RS-232 interface
- LCD display, with easy-to-read characters
- Protection against charged capacitors
- Small size and light weight
|
Data Sheet  |
|
A range of easy to use LCR meters with comprehensive measurement functions.
Accurate and fast measurements may be made from 20Hz to 1MHz (43100) making
the 4300 series the ideal choice for many applications in the test
or production environments.
All 4300 series instruments may be externally controlled using the standard RS232 or GPIB interfaces.
Technical Specifications
Measurement parameters
Any of the following parameters can be measured and displayed:
Impedance (Z) Phase Angle (Ө)
Capacitance (C) Dissipation Factor (D)
Inductance (L) Quality Factor (Q)
AC Resistance (Rac) Reactance (X)
Susceptance (B) Conductance (G)
Admittance (Y)
DC Resistance (Rdc) is an option (/R) on all models.
Series/parallel equivalent circuit
Any combination
Polar Form
Z + Ө, Y + Ө
Test conditionsFrequency range
4310 20 Hz to 100 kHz
4320 20 Hz to 200 kHz
4350 20 Hz to 500 kHz
43100 20 Hz to 1 MHz
Frequency steps
5 Hz
20 Hz to 1 kHz
50 Hz 1 kHz to 10 kHz
500 Hz 10 kHz to 100kHz
5 kHz 100 kHz to 1MHz
Drive level (AC measurements)
Open circuit: 10 mV to 2V rms
Voltage step size: 10 mV (200 steps)
Short circuit: 100 µA to 20 mA rms
Current step size: 100 µA (200 steps)
Signal source impedance: 100Ω
DC bias voltage (internal)
2 Vdc
DC bias voltage (external)
External supply of up to ±40 V may be connected via rear panel
Measurement ranges
Z, R, X 0.01 mΩ to >1 GΩ
Y, B, G 0.1 nS to >2 kS
L 1 nH to >2 kH
C 0.01 pF to >1 F
D, Q 0.0001 to 9999.9
Phase Angle -180° to +180°
Rdc 0.1 mΩ to >100 MΩ
(Option)Basic accuracy
0.1%
Varies with speed, frequency, AC drive level and impedance of Device Under Test.
Measurement speeds
Four selectable speeds for all measurement functions.
Up to 50 measurements per second depending on test frequency
and set up conditions.
Measurement connections
4 front panel BNC sockets
4-wire (Kelvin) measurements with screen at ground potential
Binning (optional)
User programmable PASS and FAIL bins.
25 way D-type interface connector.
Option /B1 (non-isolated)
Common 0V. Bin outputs 0 to 5V (nominal) with >10 mA current sink capability.
Option /B2 (isolated)
Common 24V input. Outputs 0 to 24V with >10 mA current source capability.
General
Power requirements
90 V AC to 264 V AC (Autoranging); 47 Hz to 63 Hz
Display
High contrast monochrome LCD module
320 x 240 with back lighting
Remote control
RS232 and GPIB (IEEE 488.2)
Mechanical
Height 104mm (4.1″)
Width 322mm (12.7″)
Depth 285mm (11.1″)
Weight 3kg (6.6lb)
Environmental conditions
Temperature Range
Storage -40°C to 70°C
Operating 0°C to 40°C
Full Accuracy 23°C ±5°C
Relative humidity
Up to 80% non-condensing
Altitude
Up to 2000m
Installation category
II in accordance with IEC664
Pollution degree
2 - mainly non-conductive
This equipment is intended for indoor use only in a non-explosive and
non-corrosive atmosphere
Safety
Complies with the requirements of EN61010-1:2001
EMC
Complies with EN61326 for emissions and immunity
Order codes and options
4310 LCR Meter
20 Hz to 100 kHz |
|
1J4310 |
|
4320 LCR Meter
20 Hz to 200 kHz |
|
1J4320 |
|
4350 LCR Meter
20 Hz to 500 kHz |
|
1J4350 |
|
43100 LCR Meter
20 Hz to 1 MHz |
|
1J43100 |
|
All models supplied with User Manual, Kelvin clips
and AC power cable.
|
|
Options
|
|
DC Resistance (Rdc) |
|
add /R to part number |
Bin handler - non-isolated
(cannot be fitted with /B2) |
add /B1 to part number |
Bin handler - isolated 24V
(cannot be fitted with
/B1) |
add /B2 to part number |
|
Accessories
|
| Kelvin clips (fine jaws) |
|
1EVA40150 |
| Kelvin clips (large jaws) |
1EVA40180 |
| BNC to 4-terminal component fixture
|
1EV1006 |
| 4-terminal lead set |
1EV1505 |
| SMD Probe Lead |
1EV1905A |
| SMD Tweezers |
1EVA40120 |
| Operating manual |
9H4300 |
| Certificate of Calibration with results |
1JCALRES |
|
|
|
Wayne Kerr's policy is one of continuous development and
consequently the product may vary in detail from any published
description and specification. |
| |









|